Partial volume effect correction in PET using regularized iterative deconvolution with variance control based on local topology

Kirov, A. S.*; Piao, J. Z.; Schmidtlein, C. R.
Physics in Medicine and Biology2008-5-21DOI: 10.1088/0031-9155/53/10/009
58被引1阅读
引用收藏
分享

摘要

Correcting positron emission tomography (PET) images for the partial volume effect (PVE) due to the limited resolution of PET has been a long-standing challenge. Various approaches including incorporation of the system response function in the reconstruction have been previously tested. We present a post-reconstruction PVE correction based on iterative deconvolution using a 3D maximum likelihood expectation-maximization (MLEM) algorithm. To achieve convergence we used a one step late (OSL) regularization procedure based on the assumption of local monotonic behavior of the PET signal following Alenius et al. This technique was further modified to selectively control variance depending on the local topology of the PET image. No prior 'anatomic' information is needed in this approach. An estimate of the noise properties of the image is used instead. The procedure was tested for symmetric and isotropic deconvolution functions with Gaussian shape and full width at half-maximum (FWHM) ranging from 6.31 mm to infinity. The method was applied to simulated and experimental scans of the NEMA NU 2 image quality phantom with the GE Discovery LS PET/CT scanner. The phantom contained uniform activity spheres with diameters ranging from 1 cm to 3.7 cm within uniform background. The optimal sphere activity to variance ratiowas obtained when the deconvolution function was replaced by a step function few voxels wide. In this case, the deconvolution method converged in similar to 3-5 iterations for most points on both the simulated and experimental images. For the 1 cm diameter sphere, the contrast recovery improved from 12% to 36% in the simulated and from 21% to 55% in the experimental data. Recovery coefficients between 80% and 120% were obtained for all larger spheres, except for the 13 mm diameter sphere in the simulated scan (68%). No increase in variance was observed except for a few voxels neighboring strong activity gradients and inside the largest spheres. Testing the method for patient images increased the visibility of small lesions in non-uniform background and preserved the overall image quality. Regularized iterative deconvolution with variance control based on the local properties of the PET image and on estimated image noise is a promising approach for partial volume effect corrections in PET.

学科与关键词

关键词
POSITRON-EMISSION-TOMOGRAPHYMONTE-CARLO SIMULATIONSSPATIAL-RESOLUTIONIMAGE-RECONSTRUCTIONCOMPUTED-TOMOGRAPHYVALIDATIONQUANTITATIONRANGENOISE
相关
加载中…

文献加载失败,请重试

重试

发表信息

论文状态
公开发表
期刊
Physics in Medicine and Biology
发表日期
2008-05-21
卷期
第 53 卷, 第 10 期
页码
2577-2591
DOI
10.1088/0031-9155/53/10/009
语言
MULTI-LANGUAGE
成果收录
SCIE

作者

  • K

    Kirov, A. S. (通讯) (第一作者)

  • P

    Piao, J. Z.

查看全部 3 位作者收起作者
  • S

    Schmidtlein, C. R.

登录解锁更多功能

  • 收藏这篇,随时在“我的 · 收藏”找回
  • 认领你的成果,生成推广素材
  • 关注作者获取最新成果
  • 加入项目组、协同写作
注册 / 登录

产品服务

  • 科研之友
  • 创新城
  • 科创云

服务支持

  • 帮助中心
  • 隐私政策
  • 服务条款

联系方式

在线客服:【】
客服热线:400-1616-289
电子邮箱:support@scholarmate.com

关注或下载科研之友

微信二维码
微信公众号
客户端下载二维码
下载客户端
科研成果科研人员科研机构科研动态爱瑞思软件

©2026 深圳市科研之友网络服务有限公司

公安备案图标粤公网安备 44030502000213
粤ICP备 16046710 号粤B2-20110417