摘要
Objectives: To define a dose-reduced coronary calcium CT protocol that provides similar calcium score values as a conventional 120 kVp protocol. Methods: A thorax phantom containing 100 calcifications was scanned with the reference protocol (120 kVp, 90 ref mAs, FBP) and 30 dose-reduced protocols (70-110 kVp, 90 ref mAs, FBP and iterative reconstruction (IR) levels 1-5) with 3rd generation dual-source CT. For protocols that yielded similar detectability and calcium scores as the reference protocol, additional scans were acquired at reduced ref mAs. Kendall's tau b and independent-samples-median test were used to determine trends and differences in contrast/signal-to-noise ratio (CNR and SNR), dose and calcium scores. Results: The detectability and calcium scores decreased at increasing IR levels (tau b < -0.825, p < 0.001) and increasing tube voltage (tau b < -0.679, p < 0.001). For 90 kVp-IR3 and 100 kVp-IR1, similar detectability and calcium score was found compared to the reference protocol (p > 0.206). For these protocols, lower tube currents did not affect the detectability and Agatston score (p > 0.206), while CNR and SNR were similar/higher compared to the reference protocol (0.008 < p < 0.206). Dose reduction was 60.6% (90 kVp-IR3) and 43.6% (100 kVp-IR1). Conclusions: The protocol of 90 kVp-IR3 and 100 kVp-IR1 yielded similar calcium detectability, Agatston score and image quality as the reference protocol, with dose reduction up to 60.6%.
学科与关键词
- Dose reduction for coronary calcium scoring with hybrid and model-based iterative reconstruction: an ex vivo studyden Harder, Annemarie M.; Willemink, Martin J.*; Bleys, Ronald L. A. W.; de Jong, Pim A.; Budde, Ricardo P. J.; Schilham, Arnold M. R.; Leiner, TimInternational Journal of Cardiovascular Imaging, 2014, 30(6): 1125-1133.被引 19
- Approaches to ultra-low radiation dose coronary artery calcium scoring based on 3rd generation dual-source CT: A phantom studyMcQuiston, Andrew D.; Muscogiuri, Giuseppe; Schoepf, U. Joseph*; Meinel, Felix G.; Canstein, Christian; Varga-Szemes, Akos; Cannao, Paola M.; Wichmann, Julian L.; Allmendinger, Thomas; Vliegenthart, Rozemarijn; De Cecco, Carlo N.EUROPEAN JOURNAL OF RADIOLOGY, 2016, 85(1): 39-47.被引 25
- Influence of dose reduction and iterative reconstruction on CT calcium scores: a multi-manufacturer dynamic phantom studyvan der Werf, N. R.*; Willemink, M. J.; Willems, T. P.; Greuter, M. J. W.; Leiner, T.International Journal of Cardiovascular Imaging, 2017, 33(6): 899-914.被引 13