High-Pitch Photon-Counting Detector Computed Tomography Angiography of the Aorta Intraindividual Comparison to Energy-Integrating Detector Computed Tomography at Equal Radiation Dose

Euler, Andre; Higashigaito, Kai; Mergen, Victor; Sartoretti, Thomas; Zanini, Bettina; Schmidt, Bernhard; Flohr, Thomas G.; Ulzheimer, Stefan; Eberhard, Matthias; Alkadhi, Hatem*
Investigative Radiology2022-2DOI: 10.1097/RLI.0000000000000816
100被引1阅读
引用收藏
分享

摘要

Purpose: The aims of this study were to determine the objective and subjective image quality of high-pitch computed tomography (CT) angiography of the aorta in clinical dual-source photon-counting detector CT (PCD-CT) and to compare the image quality to conventional dual-source energy-integrating detector CT (EID-CT) in the same patients at equal radiation dose. Materials andMethods: Patients with priorCTangiography of the thoracoabdominal aorta acquired on third-generation dual-source EID-CT in the high-pitchmode and with automatic tube voltage selection (ATVS, reference tube voltage 100 kV) were included. Follow-up imaging was performed on a first-generation, clinical dual-source PCD-CT scanner in the high-pitch and multienergy (QuantumPlus) mode at 120 kV using the same contrast media protocol as with EID-CT. Radiation doses between scans were matched by adapting the tube current of PCD-CT. Polychromatic images for both EID-CT and PCD-CT (called T3D) and virtual monoenergetic images at 40, 45, 50, and 55 keV for PCD-CTwere reconstructed. Computed tomography attenuation was measured in the aorta; noise was defined as the standard deviation of attenuation; contrast-to-noise ratio (CNR) was calculated. Subjective image quality (noise, vessel attenuation, vessel sharpness, and overall quality) was rated by 2 blinded, independent radiologists. Results: Forty patients were included (mean age, 63 years; 8 women; mean body mass index [BMI], 26 kg/m(2)). There was no significant difference in BMI, effective diameter, or radiation dose between scans (all P's > 0.05). The ATVS in EID-CT selected 70, 80, 90, 100, 110, and 120 kV in 2, 14, 14, 7, 2, and 1 patients, respectively. Mean CNR was 17 +/- 8 for EID-CT and 22 +/- 7, 20 +/- 6, 18 +/- 5, 16 +/- 5, and 12 +/- 4 for PCD-CT at 40, 45, 50, 55 keV, and T3D, respectively. Contrast-to-noise ratio was significantly higher for 40 and 45 keV of PCD-CT as compared with EID-CT (both P's < 0.05). The linear regression model (adjusted R-2, 0.38; P < 0.001) revealed that PCD-CT reconstruction (P < 0.001), BMI group (P = 0.007), and kVof the EID-CT scan (P = 0.01) were significantly associated with CNR difference, with an increase by 34% with PCD-CT for overweight as compared with normal weight patients. Subjective image quality reading revealed slight differences between readers for subjective vessel attenuation and sharpness, whereas subjective noise was rated significantly higher for 40 and 45 keV (P < 0.001) and overall quality similar (P > 0.05) between scans. Conclusions: High-pitch PCD-CT angiography of the aorta with VMI at 40 and 45 keV resulted in significantly increased CNR compared with EID-CT with ATVS at matched radiation dose. The CNR gain of PCD-CT increased in overweight patients. Taking into account the subjective analysis, VMI at 45 to 50 keV is proposed as the best trade-off between objective and subjective image quality.

学科与关键词

关键词
photon-counting CTmultienergyCT angiographyvirtual monoenergetic imagingaortaimage qualityradiation dose
相关
加载中…
  • Evaluation of ECG-Gated, High-Pitch Thoracoabdominal Angiographies With Dual-Source Photon-Counting Detector Computed Tomography
    Rippel, K.; Luitjens, J.; Habeeballah, O.; Scheurig-Muenkler, C.; Bette, Stefanie; Braun, Franziska; Kroencke, T. J.*; Schwarz, F.; Decker, J. A.Journal of Endovascular Therapy, 2024.被引 2
  • Impact of CT Photon-Counting Virtual Monoenergetic Imaging on Visualization of Abdominal Arterial Vessels
    Dillinger, Daniel*; Overhoff, Daniel; Booz, Christian; Kaatsch, Hanns L.; Piechotka, Joel; Hagen, Achim; Froelich, Matthias F.; Vogl, Thomas J.; Waldeck, StephanDiagnostics, 2023, 13(5): 938.被引 21
  • Individualized CT Angiography Protocols for the Evaluation of the Aorta: A Feasibility Study
    Kok, Madeleine; de Haan, Michael W.; Mihl, Casper; Eijsvoogel, Nienke G.; Hendriks, Babs M. F.; Sailer, Anna M.; Derks, Kris; Schnerr, Roald S.; Schurink, Geert Willem H.; Wildberger, Joachim E.; Das, Marco*Journal of Vascular and Interventional Radiology, 2016, 27(4): 531-538.被引 11
查看全部 →

发表信息

论文状态
公开发表
期刊
Investigative Radiology
发表日期
2022-02
卷期
第 57 卷, 第 2 期
页码
115-121
DOI
10.1097/RLI.0000000000000816
语言
ENGLISH
期刊等级
1区
成果收录
SCIE

作者

  • E

    Euler, Andre (第一作者)

  • H

    Higashigaito, Kai

查看全部 10 位作者收起作者
  • M

    Mergen, Victor

  • S

    Sartoretti, Thomas

  • Z

    Zanini, Bettina

  • S

    Schmidt, Bernhard

  • F

    Flohr, Thomas G.

  • U

    Ulzheimer, Stefan

  • E

    Eberhard, Matthias

  • A

    Alkadhi, Hatem (通讯)

登录解锁更多功能

  • 收藏这篇,随时在“我的 · 收藏”找回
  • 认领你的成果,生成推广素材
  • 关注作者获取最新成果
  • 加入项目组、协同写作
注册 / 登录

产品服务

  • 科研之友
  • 创新城
  • 科创云

服务支持

  • 帮助中心
  • 隐私政策
  • 服务条款

联系方式

在线客服:【】
客服热线:400-1616-289
电子邮箱:support@scholarmate.com

关注或下载科研之友

微信二维码
微信公众号
客户端下载二维码
下载客户端
科研成果科研人员科研机构科研动态爱瑞思软件

©2026 深圳市科研之友网络服务有限公司

公安备案图标粤公网安备 44030502000213
粤ICP备 16046710 号粤B2-20110417