Ultra-low-dose spectral-detector computed tomography for the accurate quantification of pulmonary nodules: an anthropomorphic chest phantom study

Zhou, Xiuxiu; Zhang, Hanxiao; Jin, Xiaoxing; Zhang, Xiaohui; Lu, Xiaomei; Han, Qun; Xiong, Xiaoge; Liu, Tian; Feng, Yan; Tu, Wenting; Zhou, Taohu; Ge, Yanming; Dong, Peng; Liu, Shiyuan*; Fan, Li*xuzhou med univ; shanghai univ finance & econ; weifang med univ
Diagnostic and Interventional Radiology2023-9DOI: 10.4274/dir.2023.232233
5阅读
引用收藏
分享

摘要

PURPOSE To assess the quantification accuracy of pulmonary nodules using virtual monoenergetic images ( VMIs) derived from spectral-detector computed tomography (CT) under an ultra-low-dose scan protocol. METHODS A chest phantom consisting of 12 pulmonary nodules was scanned using spectral-detector CT at 100 kVp/10 mAs, 100 kVp/20 mAs, 120 kVp/10 mAs, and 120 kVp/30 mAs. Each scanning protocol was repeated three times. Each CT scan was reconstructed utilizing filtered back projection, hybrid iterative reconstruction, iterative model reconstruction (IMR), and VMIs of 40-100 keV. The signal-to-noise ratio and air noise of images, absolute differences, and absolute percentage measurement errors (APEs) of the diameter, density, and volume of the four scan protocols and ten reconstruction images were compared. RESULTS With each fixed reconstruction image, the four scanning protocols exhibited no significant differences in APEs for diameter and density (all P > 0.05). Of the four scan protocols and ten reconstruction images, APEs for nodule volume had no significant differences (all P > 0.05). At 100 kVp/10 mAs, APEs for density using IMR were the lowest (APE(-mean): 6.69), but no significant difference was detected between VMIs at 50 keV (APE(-mean): 11.69) and IMR (P = 0.666). In the subgroup analysis, at 100 kVp/10 mAs, there were no significant differences between VMIs at 50 keV and IMR in diameter and density (all P > 0.05). The radiation dose at 100 kVp/10 mAs was reduced by 77.8% compared with that at 120 kVp/30 mAs. CONCLUSION Compared with IMR, reconstruction at 100 kVp/10 mAs and 50 keV provides a more accurate quantification of pulmonary nodules, and the radiation dose is reduced by 77.8% compared with that at 120 kVp/30 mAs, demonstrating great potential for ultra-low-dose spectral-detector CT.

学科与关键词

关键词
Spectral computed tomographychest phantompulmonary nodulelow-dose computed tomographyreconstruction algorithm
相关
加载中…
  • Ultra-low dose dual-layer detector spectral CT for pulmonary nodule screening: image quality and diagnostic performance
    Ding, Li; Chen, Mingwang; Li, Xiaomei; Wu, Yuting; Li, Jingxu; Deng, Shuting; Xu, Yikai*; Chen, Zhao*; Yan, Chenggong*Insights into Imaging, 2025, 16(1): 11.
  • Prospective Study of Low- and Standard-dose Chest CT for Pulmonary Nodule Detection: A Comparison of Image Quality, Size Measurements and Radiation Exposure
    Hu, Qiong-jie; Liu, Yi-wen; Chen, Chong; Kang, Shi-chao; Sun, Zi-yan; Wang, Yu-jin; Xiang, Min; Xia, Li-ming; Guan, Han-xiong*Current Medical Science, 2021, 41(5): 966-973.被引 3
  • Ultra-low dose computed tomography protocols using spectral shaping for lung cancer screening: Comparison with low-dose for volumetric LungRADS classification
    Milanese, Gianluca; Ledda, Roberta Eufrasia; Sabia, Federica; Ruggirello, Margherita; Sestini, Stefano; Silva, Mario; Sverzellati, Nicola; Marchiano, Alfonso Vittorio; Pastorino, Ugo*EUROPEAN JOURNAL OF RADIOLOGY, 2023, 161: 110760.被引 9
查看全部 →

发表信息

论文状态
公开发表
期刊
Diagnostic and Interventional Radiology
发表日期
2023-09
卷期
第 29 卷, 第 5 期
页码
691-703
DOI
10.4274/dir.2023.232233
语言
ENGLISH
成果收录
SCIE
资助标注
National Natural Science Foundation of China [81871321, 82171926, 81930049, 82202140]; program of Science and Technology Commission of Shanghai Municipality [21DZ2202600]; National Key R&D Program of China [2022YFC2010000, 2022YFC2010002, 2017YFC1308703]; construction of CT standardized database for chronic obstructive pulmonary disease [YXFSC2022JJSJ002]; Science and Technology Innovation Action Plan Project of Shanghai [19411951300]; Shanghai Sailing Program [20YF1449000]; Clinical Innovation Project of Shanghai Changzheng Hospital [2020YLCYJ-Y24]

作者

  • Z

    Zhou, Xiuxiu (第一作者)

  • Z

    Zhang, Hanxiao

    xuzhou med univ

查看全部 15 位作者收起作者
  • J

    Jin, Xiaoxing

  • Z

    Zhang, Xiaohui

  • L

    Lu, Xiaomei

  • H

    Han, Qun

  • X

    Xiong, Xiaoge

    shanghai univ finance & econ

  • L

    Liu, Tian

  • F

    Feng, Yan

  • T

    Tu, Wenting

  • Z

    Zhou, Taohu

    weifang med univ

  • G

    Ge, Yanming

    weifang med univ

  • D

    Dong, Peng

    weifang med univ

  • L

    Liu, Shiyuan (通讯)

  • F

    Fan, Li (通讯)

登录解锁更多功能

  • 收藏这篇,随时在“我的 · 收藏”找回
  • 认领你的成果,生成推广素材
  • 关注作者获取最新成果
  • 加入项目组、协同写作
注册 / 登录

产品服务

  • 科研之友
  • 创新城
  • 科创云

服务支持

  • 帮助中心
  • 隐私政策
  • 服务条款

联系方式

在线客服:【】
客服热线:400-1616-289
电子邮箱:support@scholarmate.com

关注或下载科研之友

微信二维码
微信公众号
客户端下载二维码
下载客户端
科研成果科研人员科研机构科研动态爱瑞思软件

©2026 深圳市科研之友网络服务有限公司

公安备案图标粤公网安备 44030502000213
粤ICP备 16046710 号粤B2-20110417